OCMA-500

Oil Content Analyzer

OCMA-500 is an Oil Content Analyzer designed for accurate oil content measurement in wastewater and environmental water. This oil content analyzer enables automatic measurement through simple one-touch operation and is suitable for applications such as wastewater monitoring and component quality control where reliable oil concentration measurement is required.

OCMA-500

Why Choose OCMA-500

  • One-touch automatic operation enables easy oil content measurement by anyone using this oil content analyzer
  • Solvent extraction and non-dispersive infrared absorption method supports reliable oil concentration analysis
  • Fast measurement process significantly reduces analysis time compared to conventional oil measurement methods
  • The oil content analyzer complies with ASTM D7066-4 for oil and grease determination in water
Why Choose

Key Advantages of OCMA-500

1.

Automatic Oil Content Measurement

Single button

This oil content analyzer performs automatic measurement from oil extraction to result display with minimal operator involvement.

2.

Rapid Oil Concentration Analysis

Approx. 3 minutes

Quick oil content measurement enables faster analysis compared to traditional oil analysis methods.

3.

Wide Oil Measurement Range

0–200 mg/L

The oil content analyzer supports oil concentration measurement across a wide range without changing measurement conditions.

4.

Low-Boiling Oil Measurement Capability

No evaporation

By eliminating solvent evaporation, this oil content analyzer prevents loss of low-boiling oil components during oil measurement.

Applications of OCMA-500

Wastewater Oil Monitoring

Oil content measurement in factory drainage and environmental water prior to discharge.

Factory Drainage Oil Analysis

Oil concentration measurement in industrial drainage, including samples containing surfactants with pre-treatment.

Environmental Water Oil Measurement

Use of the oil content analyzer for oil concentration analysis in environmental water monitoring activities.

Component Oil Contamination Control

Oil content measurement for evaluating residual oil during component manufacturing and quality control processes.

Technical Specifications of OCMA-500

Product name Oil content analyzer
Model OCMA-500
Measurement method Solvent extraction – non-dispersive infrared absorption analysis method
Measured objects Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm
Measurement range 0 mg/L to 200 mg/L
Resolution For mg/L 0 to 99.9: 0.1; 100 to 200: 1
Repeatability 0 mg/L to 9.9 mg/L: ±0.3 mg/L
10.0 mg/L to 99.9 mg/L: ±2.1 mg/L
100 mg/L to 200 mg/L: ±5 mg/L
Display 3.5 inches, 320×240 dots, backlight color graphic LCD
Calibration method Zero and span calibration
Amount of test sample required Sample water : Solvent = 2:1
Extraction solvent S-316
Amount of extraction solvent required 8 mL
Extraction method Built-in extractor
Ambient operating temperature 0°C to 40°C (no condensation)
Power supply AC 100 V to 240 V ±10%, 50/60 Hz
Power consumption AC 100 V: Approx. 60 VA; AC 240 V: Approx. 90 VA
External dimensions 342 (H) × 200 (W) × 313 (D) mm
Mass Approx. 7 kg
External output Output to a USB memory stick
Functions 300-item data memory (measurement history)
Self error determination
Stabilized measurement value display
Clock function
Unit conversion function
Compliant with ASTM D7066-4

NOT SURE WHAT YOU ARE LOOKING FOR?

BOOK A CONSULTATION

Our experts are here to guide you to the right solution. Fill out the form and we'll get in touch shortly.

*
*
*
*
*
*
India
*
Type or select city
*
Select region
*
Sales
*
Select your segment
*

I agree to be contacted by HORIBA India/partners via call, SMS, WhatsApp, and email, and accept the Terms & Conditions and Privacy Policy

Frequently Asked Questions

OCMA-500 measures oil content extracted from wastewater and environmental water using a solvent extraction and non-dispersive infrared absorption method.

The oil content analyzer supports oil concentration measurement from 0 to 200 mg/L.

Oil content measurement can be completed in approximately 3 minutes by pressing the start button, excluding warm-up and calibration time.

Yes, the OCMA-500 Oil Content Analyzer complies with ASTM D7066-4 for oil and grease determination in water.

The oil content analyzer is used for measuring oil extracted from wastewater, factory drainage, and environmental water, and can also be applied for component quality control.

Yes, the OCMA-500 does not require solvent evaporation, which helps prevent volatilization of low-boiling oil components during oil content measurement.

Yes, the oil content analyzer features one-touch automatic operation, enabling oil content measurement by simply pressing a single button.