TX-100

TX-100

Tunable Laser Gas Analyzer

The TX-100 is an in‑situ, probe‑type laser-based HCl analyzer for continuous, real‑time measurement of hydrogen chloride concentrations in flue gas streams. Its compact single‑flange installation and continuous moisture correction ensure high‑accuracy HCl control in incineration plants, power generation, and chemical processing.

Why Choose TX-100

  1. Probe‑type optical system for in‑situ HCl measurement
  2. Both laser and receiver on same‑side probe for compact operation
  3. Single‑flange installation reduces setup complexity
  4. Easy swap‑in replacement with existing analyzers
  5. Stabilized optical axis ensures measurement precision
  6. Continuous moisture correction maintains accuracy in humid gases
Why Choose

Key Advantages of TX-100

Simplified installation 

One‑flange probe mount

Minimizes installation complexity in tight stacks

Robust measurement stability 

Stabilized optical axis

Ensures repeatable HCl readings over long‑term operation

Precise moisture compensation

Continuous moisture correction

 Maintains accuracy in high‑humidity flue gases

Compact probe design 

Laser and receiver on same side

Saves space and simplifies maintenance access

Technical Specifications of TX-100

NameTunable Laser Gas Analyzer
ModelTX-100
Measurement targetHCl and H2O in process gas
Measurement componentHCl, or HCl and H2O simultaneous measurement
Measurement principleNon Dispersive Infrared (NDIR)
Measurement typeProbe type
Laser classCLASS1(IEC60825-1: 2014)
Protection degreeIP65
Wetted materialsSUS316L、SUS304、BK7、FKM、PTFE、aluminum alloy
Tube connectionsRc1/8 (Tube of 8 mm O.D./6 mm I.D.)
External dimentionsAnalyzer: Max: 358 (W) x 517 (D) x 453 (H) mm
Probe
Flange diameter (Max.):  Φ235 mm、 Length: 1500 mm
Probe diameter: Φ 89 mm
WeightAnalyzer: 19 kg
Probe: 20 kg
Flange specificationsJIS10K-100AFF、JIS5K-125AFF、ANSI 150 Lb 4BRF
Power supply voltageDC24V±10%
Power consumption90 W
DisplayLCD (128 × 64 dots), LED (3 units: Status display)
Display contentsGas concentration, transmission, process gas temperature, process gas pressure, system status, alarm information, system information (product serial no., output signal, IP address, optical path length, analyzer unit internal temperature)
CommunicationModbusTM/TCP
Analog input/outputInput:
Two channels (Process gas temperature, process gas pressure)
4 to 20 mA (isolated input)
Output:
Two channels (Process gas concentration*, transmission, process gas temperature, process gas pressure)
4 to 20 mA (isolated output) Max. load: 550 Ω
Digital input/outputInput:
One channel (External alarm) 
Non-voltage digital input (isolated input) 
Input open signal: 100 kΩ or more 
Input close signal: 200 Ω or less
Output:
Two channels (Warning/Calibration/Warm-up/Maintenance state/Fault)
Relay contact output (isolated output) Max. 24 V, 1 A at each terminal

*You can select either fixed value or real time HCl correction using H2O concentration.
-Modbus is a trademark of Schneider Electric Inc.
-Consult HORIBA if special probe length is required.

 

Performance

Measurement componentHCl, H2O
Measurement range (250℃、101.3kPa)

HCl: 0-50,100,200,300,400,500 ppm*1

H2O: 0-50 vol%

Measured optical path length2m
Repeatability

HCl: +/- 1.0% F.S.

H2O: +/- 2.0% F.S.

Linearity

HCl: +/- 1.0% F.S.

H2O: +/- 2.0% F.S.

Zero drift+/- 2.0% F.S. / 6 months
Span drift+/- 2.0% F.S. / 6 months
Data update cycleApprox. 2sec
Recommended Calibration cycle6 months
Calibration methodHCl: Manual calibration using gas cell
H2O: Factory calibration*2
Interference+/- 2.0% F.S.
Gas conditions (at standard process gas condition range)
Warm-up time3hours

*1: Select a measurement range.
*2: You can adjust the H2O indicated value to the manual analysis value. Consult HORIBA for more details. 

Usage Conditions

Ambient temperature-20 ~ 55°C (no direct sun rays, radiant heat and heat conduction from flange)
Ambient humidityRelative humidity: 90% or less (no condensation)
DustAmbient criterion or less
Air purgeInstrument air (must not contain oil or mist; no dew formation)
Pressure: 0.2 to 0.25 MPa*
Air purge flow rate60 L/min or more (depend on the process gas conditions)
VibrationAt 1 to 200 Hz, 1.96 m/s2 or less (excluding resonant frequency)
Process gas conditions

Temperature: 150 to 250°C

Pressure: 101.3 ± 10 kPa

Flow speed: 25m/s

Dust: 0.1mg/Nm3 or less

* This value is supplied pressure to TX-100 main unit.  In case air passes through the purge unit, the pressure value at the purge unit is 0.2 to 0.9 MPa.

Related Products

IT-270

IT-270

High-Accuracy Infrared Thermometer

ENDA-5000 Series

ENDA-5000 Series

Stack Gas Analysis System

APMA-370

APMA-370

Ambient Carbon Monoxide Analyzer

APOA-370

APOA-370

Ambient Ozone Monitor

GA-370

GA-370

Trace Gas Monitor

PLGA-1000

PLGA-1000

Process Laser Gas Analyzer

NOT SURE WHAT YOU ARE LOOKING FOR?

BOOK A CONSULTATION

Our experts are here to guide you to the right solution. Fill out the form and we'll get in touch shortly.

*
*
*
*
*